Length calibration

50 meter tape benchNWML maintains world-class laser-based length calibration benches which are housed in purpose-built laboratories. The facilities are based on Hewlett-Packard helium-neon laser interferometers and custom-built track and carriage systems. This service has full UKAS accreditation for the calibration of line standards and measures up to 50 metres and above in 50 m multiples from the lowest to the highest quality.

Range of value tested

Best measurement Capability

Accredited

Rigid line measures

0 – 2.3 metres

(1 + 1L) µm

where L is in metres

UKAS

Flexible line measures 2 - 50 metres

(10 +0.6L) µm

where L is in metres

UKAS

Rigid line standards can be calibrated by direct comparison with the laser wavelength. The mounting for the standard is of a design to allow for all the commonly met cross-sections (H, flat & bevelled).

Flexible line standards can be calibrated either by comparison with NWML standard tapes or by direct comparison with a laser.

For free advice about length calibration or a price quotation, please contact Bob Hynds or Robert Harper on +44(0)20 8943 7272 or e-mail bob.hynds@nmo.gov.uk or robert.harper@nmo.gov.uk